Investigators can find current ascii tables behind the following links containing the Kepler
instrument response curve in hi-res and
low-res tabulations.
Figure 1: The Kepler focal plane detectors and the optical elements within the Kepler telscope.
From the Kepler Instrument Handbook (KIH).
The primary instrument aboard Kepler is the focal plane array
consisting of 21 science and 4 Fine Guidance Sensor CCD
modules. Field flattener lenses on each module map the spherical
telescope image surface onto the flat CCD chips, and define the
overall wavelength bandpass. Each science module is an array
of 2200 by 2048 pixels. These 21 modules each have 4 output
channels, for a total of 84 channels and 94.6 million active
pixels that view the sky, with additional masked real pixels and
virtual pixels for collection of collateral data.
The shape of the bandpass, described below, was chosen to contain
most of the optical spectrum. This choice maximizes the
sensitivity of the telescope / detector combination for detecting
planets transiting solar-type stars. Kepler contains
no "true" filter, in the sense that HST imaging
instruments include filter wheel assemblies with multiple,
specifically-defined bandpasses. Here, the intent is to utilize the
entire optical range, except for the short wavelengths, which
were truncated to avoid chromspheric emission lines in solar-type
stars. The photometer provides no color information, but does
provide excellent depth. Kepler's wide-band images are similar to
the clear filter frames taken with HST/STIS.
To achieve maximum sensitivity, the Kepler bandpass is wider than the
typical broad-band filters commonly used in optical astronomy
(e.g. Johnson UBVRI, Sloan ugriz). The absolute Kepler
sensitivity curve is displayed in the clickable figure (Fig 2) to the right. The displayed response
curve was derived during pre-flight testing, and represents the
laboratory calibration of the Kepler photometer. It is this curve which
is provided in the ascii tables at the top of the page in both high- and low-resolution forms.
Figure 3: Optical element components of the Kepler Instrument Response compared
to approximate M5 and G2 stellar spectra.
The total photometer spectral response is a combination of the
transmission functions of all optical elements, including the Schmidt
corrector, the primary mirror assembly, the field flatterner lenses
on each CCD module, and the wavelength dependent quantum efficiency
of the detectors (Fig 3). The front surfaces of the field flatteners are
anti-reflection coated; a bandpass filter coating was applied to the back
surfaces. This bandpass was chosen to minimize the effects of stellar
variability in the near-ultraviolet (λ ≤ 420 nm), especially the
Ca II H & K emission lines, which would impact exooplanet transit
detection. At long wavelengths the coating was designed to minimize fringing.
Fig 4: Comparison of the Kepler, MOST, CoRoT and Johnson response curves. Kindly provided
by Jason Rowe (NASA Ames) and extracted from
2009IAUS..253..121R.
Fig 4 compares the Kepler response function with those of two similar
missions: the
MOST spacecraft, and the CoRoT
mission. It shows the response functions
from 400 to 900 nm. The MOST bandpass bandpass is marked
by the dashed line, the Kepler bandpass is shown in black and the
CoRoT whitelight bandpass is shown by the dot-dashed line. The
transmission functions for the Johnson B,V,R,I filters are shown from
left to right in blue, green, red and magenta respectively and have
been scaled to peak at 100% transmission. The spectrum for an A2V
star is shown in cyan, which peaks in the UV and the spectrum for a
M2V star is shown in orange which peaks in the infrared. The two
spectra have been scaled to have equal flux in the Johnson V
filter.
If you have a question about the Kepler GO program, please contact us via the
address.